Making Measurements
Measuring Jitter Transfer
HOW TO: Start the Jitter Transfer Measurement
NOTE The Jitter Transfer measurement must be started within 10 minutes of
the completion of calibration.
1 After the CALIBRATION is completed, remove the back to back
connection from the DSn or SONET or optical interfaces.
If the measurement is to be made at a network equipment monitor
point, choose on the DSn or SONET display
before making the jitter transfer measurement.
2 Choose MODE on the display and
press .
The measurement’s progress can be monitored on the
display.Atthe endof thetestthe graphcan beviewedon the
display.
MONITOR
RECEIVE
MEASURE
TRANSMIT
JITTER
RUN/STOP
TRANSMIT
RESULTS
JITTER