HP 8517B S-Parameter Test Set Manual
3-16
CAUTION
Important! The test set RF connector becomes loosened easily. Hold the RF
cable securely throughout the remainder of this procedure. Do not allow the
cable to rotate.
Figure 3-5 Using a Torque Wrench Correctly
Positioning the thumb-screw
See Figure 3-6. Position the clamp so the thumb-screw is positioned at the
top of the clamp.
Turn the clamp to visually align the clamp flats with the flats on the test
port connector nut.
This positioning minimizes rotating the connector in the procedure step.