HP Saw E4419B Manual de usuario Pagina 169

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 204
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 168
HP4402.book : chapter4.fm 3 Fri Feb 12 15:32:45 1999
Error Messages
Introduction
HP E4419B User’s Guide 4-3
Errors are retrieved in a first in first out (FIFO) order. If more than 30
errors occur, the error queue overflows and the last error in the queue is
replaced with error -350, “Queue Overflow”. Any time the queue
overflows the most recent error is discarded.
Once the errors are read they are removed from the error queue. This
opens a position at the end of the queue for a new error message, if one is
subsequently detected. When all errors have been read from the queue,
further error queries return +0, “No errors”.
To delete all the errors in the queue from the front panel press:
, , then .
To delete all the errors in the queue remotely use:
the *CLS (clear status) command.
The error queue is also cleared when the instrument power has been
switched off.
System
Inputs
More
Error List Clear Errors
Vista de pagina 168
1 2 ... 164 165 166 167 168 169 170 171 172 173 174 ... 203 204

Comentarios a estos manuales

Sin comentarios