HP4402.book : chapter4.fm 7 Fri Feb 12 15:32:45 1999
Error Messages
Error Messages
HP E4419B User’s Guide 4-7
-214 Trigger deadlock
TRIG:SOUR was set to HOLD or BUS and a READ? or MEASure? was
attempted, expecting TRIG:SOUR to be set to IMMediate.
-220 Parameter error;Frequency list must be in ascending order.
Indicates that the frequencies entered using the
MEMory:TABLe:FREQuency command are not in ascending order.
-221 Settings conflict
This command occurs under a variety of conflicting conditions. The
following list gives a few examples of where this error may occur:
•If the READ? parameters do not match the current settings.
•If you are in fast mode and attempting to switch on for example,
averaging, duty cycle or limits.
•Trying to clear a sensor calibration table when none is selected.
-221 Settings conflict;DTR/DSR not available on RS422
DTR/DSR is only available on the RS232 interface.
-222 Data out of range
A numeric parameter value is outside the valid range for the command.
For example, SENS:FREQ 2KHZ.
-224 Illegal parameter value
A discrete parameter was received which was not a valid choice for the
command. You may have used an invalid parameter choice.
For example, TRIG:SOUR EXT.
-226 Lists not same length
This occurs when SENSe:CORRection:CSET[1]|CSET2:STATe is set
to ON and the frequency and calibration/offset lists do not correspond in
length.
-230 Data corrupt or stale
This occurs when a FETC? is attempted and either a reset has been
received or the power meter state has changed such that the current
measurement is invalidated (for example, a change of frequency setting
or triggering conditions).