HP 181A Manual de usuario Pagina 86

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 254
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 85
Chapter 1 Performance Tests
Test 5: Option 015/030/050/124 Channel 2 Frequency Accuracy (HP
53181A Only)
1-64 Assembly-Level Service Guide
1
Procedure
1 Connect equipment as shown in Figure 1-15.
Figure 1-15. Option 015/030/050/124 Channel 2 Frequency Accuracy Test Setup
2 Set the HP 8663A Synthesized Signal Generator to output
100 MHz at 17.0 dBm.
Verify the Counter reads 100 MHz ±70 mHz.
HP 53181A
Counter
Channel 2
50
Feedthrough
Ref. In
(rear panel)
HP 8663A
Synthesized Signal Generator
N-to-BNC
Connector
Output
Input
(rear panel)
HP 8340B
Sweep Oscillator
Frequency Standard
(switch set to EXT)
Output
Vista de pagina 85
1 2 ... 81 82 83 84 85 86 87 88 89 90 91 ... 253 254

Comentarios a estos manuales

Sin comentarios