HP 132A Manual de usuario Pagina 179

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 330
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 178
Chapter 4 Command Reference
Introduction
Programming Guide 4-13
:CALCulate2:LIMit:CLEar:AUTO <Boolean>
Sets or queries if the limit test results are to be cleared with each
:INITiate[:IMMediate] and :INITiate:CONTinuous ON operation.
Single ASCII-encoded byte, 0 or 1.
A value of 0 indicates OFF; a value of 1 indicates ON.
*RST: ON
When AUTO is ON, the Counter will perform the following whenever
:INIT[:IMM] or :INIT:CONT ON is executed:
– Invalidate the limit data.
– Clear :CALC2:LIM:FAIL, :CALC2:LIM:FCOunt, and
:CALC2:LIM:PCOunt information.
– Turn off the front-panel displays Limit annunciator.
– Set the Limit-Detect output of the RS-232 connector to the
in-limit voltage level.
When AUTO is OFF, the only way to clear the limit-test results is to send
:CALC2:LIM:CLE[:IMM].
:CALCulate2:LIMit:CLEar[:IMMediate]
This command is an event that causes the Counter to
immediately invalidate the limit data,
clear the information in :CALC2:LIM:FAIL, :CALC2:LIM:FCOunt, and
:CALC2:LIM:PCOunt,
turn off the front-panel displays Limit annunciator, and
set the Limit-Detect output to its in-limit voltage level.
If :CALC2:LIM:STAT is OFF, error -221 is generated.
Query Response
Comments
Comments
Vista de pagina 178
1 2 ... 174 175 176 177 178 179 180 181 182 183 184 ... 329 330

Comentarios a estos manuales

Sin comentarios