Chapter 4 Command Reference
Introduction
Programming Guide 4-43
:HCOPy Subsystem
:HCOPy:CONTinuous <Boolean>
Enables or disables printing results.
When :HCOPy:CONTinuous is enabled (:HCOP:CONT ON),
the Counter prints each measurement.
If statistics is enabled (:CALC3:AVER[:STAT] ON), all statistics (standard
deviation, mean, minimum, and maximum ) will be printed in addition to the
individual measurements. If limit testing is enabled (:CALC2:LIMit:STAT ON),
an indication will be printed for the measurements that fail the limit test.
Refer to the sections titled “Using the Print Menu,and To Configure the RS-
232 Serial Port for Printingin the Operating Guide for more details on printing.
Single ASCII-encoded byte, 0 or 1.
A value of 0 indicates OFF, a value of 1 indicates ON.
*RST: OFF
Save & Print
Query Response
Comments
Related
Front-Panel
Keys