HP 132A Manual de usuario Pagina 251

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 330
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 250
Chapter 4 Command Reference
Introduction
Programming Guide 4-85
[:SENSe]:EVENt3:SLOPe?
Queries which edge of channel 3 input port will be considered an event.
A sequence of ASCII-encoded bytes: POS
[:SENSe]:FREQuency Subtree
This subtree controls the Frequency, Frequency Ratio, and Period measuring
capabilities of the instrument.
[:SENSe]:FREQuency:ARM Subtree
This subtree is used to synchronize the Frequency, Frequency Ratio, and Period
start and stop arm with events. The following combination of start/stop arming
sources are valid:
STARt:SOURce STOP:SOURce Front-Panel Gating Settings
GATE STOP
IMMediate IMMediate AUTO _______
IMMediate TIMer TIME _______
IMMediate DIGits DIGITS _______
EXTernal EXTernal EXTERNAL POS, NEG
EXTernal TIMer EXTERNAL TIME
EXTernal IMMediate EXTERNAL AUTO
[:SENSe]:FREQuency:ARM[:STARt]:SLOPe POSitive | NEGative
Sets or queries the slope of the external start arm signal used in external arming
Frequency, Frequency Ratio, and Period measurements.
A sequence of ASCII-encoded bytes: POS or NEG
*RST: POSitive
Only applies when [:SENS]:FREQ:ARM[:STAR]:SOUR EXT is selected.
Gate & ExtArm
Query Response
Query Response
Comments
Related
Front-Panel
Keys
Vista de pagina 250
1 2 ... 246 247 248 249 250 251 252 253 254 255 256 ... 329 330

Comentarios a estos manuales

Sin comentarios