106 Multimeter Command Reference Chapter 3
:PERiod?
MEASure:PERiod? [<range>|MIN|MAX|DEF|AUTO
[,<
resolution>|MIN|MAX|DEF]] selects the period function and allows you to specify
range and resolution.
Parameters
Comments The period function uses one “range” for all inputs between 0.33 seconds and
3.3
µ
Sec. A period measurement will return “
0” if no input is applied.
Range and resolution settings are listed below for the MIN, MAX, DEF and
AUTO parameters and after a module reset (*RST).
Parameter Name Parameter Type Range of Values Default Units
<range>
numeric 3.33E-01 Sec
<resolution>
numeric 3.33E-05| 3.33E-06 | 3.33E-07 Sec
PARAMETER RANGE RESOLUTION
MIN 3.33E-01 3.33E-07
MAX 3.33E-01 3.33E-05
DEF|AUTO
and module
reset (
*RST)
3.33E-01 3.33E-06