120 Multimeter Command Reference Chapter 3
CURRent:AC:RANGe:AUTO
[SENSe:]CURRent:AC:RANGe:AUTO <mode> enables or disables the autorange
function for AC current measurements.
Parameters
Comments You can substitute decimal values for the OFF (“0”) and ON (“1”) parameters.
When autoranging is ON, the multimeter samples the input before each
measurement and selects the appropriate range.
If you explicitly select a range using CURRent:AC:RANGe, autoranging is
turned
OFF.
Related Commands: CONFigure, :DC:RANGe, RESistance:RANGe
*RST Condition: CURR:AC:RANG:AUTO ON
Example Disable AC Current Autoranging
CURR:AC:RANG:AUTO OFF
Disable autorange.
CURRent:AC:RANGe:AUTO?
[SENSe:]CURRent:AC:RANGe:AUTO? returns a number to show whether the AC
current autorange mode is enabled or disabled:
1” = ON, “0” = OFF. The number
is sent to the output buffer.
Example Query the AC Current Autorange Mode
CURR:AC:RANG:AUTO OFF
Disable autorange.
CURR:AC:RANG:AUTO?
Query multimeter to return autorange mode.
enter statement
Enter value into computer.
Parameter Name Parameter Type Range of Values Default Units
<
mode>
boolean OFF|0|ON|1 none