HP Waterskis E1312A Manual de usuario Pagina 82

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 212
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 81
82 Multimeter Command Reference Chapter 3
Example Enter Calibration Information to Record the Next Calibration Date
CAL:STR Cal 4/4/YY, Due 10/4/YY’
Enter a calibration message to record the cal
date of April 4 and next cal due date as
October 4 (YY = year of due date).
:STRing?
CALibration:STRing? queries the calibration message and returns a quoted string
(or a null string “ ” if nothing is present).
Example Query the Calibration Message
CAL:STR?
Query the calibration message.
enter statement
Enter value into computer.
:VALue
CALibration:VALue <cal_value> specifies the value of the known calibration
signal used by the calibration procedure. See the HP E1312A and HP E1412A
Service Manual, Chapter 5 “Adjustments”, for a more detailed description of the
multimeter's calibration/adjustment procedures.
Parameters
Comment *RST does not affect the calibration value.
Example Enter the Known Value for the Calibration Source Signal
CAL:VAL 10.0
Enter calibration value.
:VALue?
CALibration:VALue? queries the present calibration value.
Example Query the Calibration Value
CAL:VAL?
Query the calibration value.
enter statement
Enter value into computer.
Parameter Name Parameter Type Range of Values Default Units
<
cal_value>
numeric See the service manual none
Vista de pagina 81
1 2 ... 77 78 79 80 81 82 83 84 85 86 87 ... 211 212

Comentarios a estos manuales

Sin comentarios