User's Guide Version 3.12
March 2015
The OSPL90 curve, the HFA (or SPA) and peak value of this curve, the frequency of the peak, the HFA (or
SPA) full-on gain are determined at full-on gain setting. The response curve, frequency range, Reference Test
Gain, battery drain, equivalent input noise, distortion, attack and release times are determined at RTS.
These changes result in more consistent values for reference test gain, equivalent input noise and attack and
release time. See 1996 or 2003 ANSI standard, 8.2: ANSI S3.22-1996 facts and 8.5: ANSI 2003 Linear and AGC
tests.
ANSI 1996 Linear and AGC tests
1. Press <Tests>, then highlight and <PICK> [Test box measures], then highlight and <PICK> [Linear] or
[AGC].
2. Highlight and <PICK> [Start test] and follow the test setup instructions which are presented.
3. Refer to Test Box MeasuresSetup, Coupling the hearing instrument, Positioning the hearing instrument and
Battery pill use and care to prepare the instrument for testing.
4. Turn the instrument ON.
5. Set the gain control (manual or programmable) to maximum with all other controls set to provide the widest
frequency response, greatest gain and highest maximum output. AGC controls and/or programmable
parameters should be set as designated by the manufacturer.
6. Press <Continue> to run the test or <Cancel> to abort the test. Note that, if required, the RM500SL will
pause during the test to request that the volume control be adjusted to the Reference Test Position (RTP).
Open the test box if necessary and, without moving the hearing instrument, adjust the user gain control until
the large vertical arrow on the screen prompt disappears. Press <Continue> to resume the test.
If you are unsure which test sequence to run, try [AGC] first. If attack and release times are all near zero, the
instrument is probably linear.
ANSI 2003 Linear and AGC tests
1. Press <Tests>, then highlight and <PICK> [Test box measures], then highlight and <PICK> [Linear] or
[AGC].
2. Highlight and <PICK> [Start test] and follow the test setup instructions which are presented.
3. Refer to Test Box MeasuresSetup, Coupling the hearing instrument, Positioning the hearing instrument and
Battery pill use and care to prepare the instrument for testing.
4. Turn the instrument ON.
5. Set the gain control (manual or programmable) to maximum with all other controls set to provide the widest
frequency response, greatest gain and highest maximum output. AGC controls and/or programmable
parameters should be set to minimize AGC action (AGC off or threshold high, compression ratio close to 1).
6. Press <Continue> to run the test or press <Cancel> to abort the test. Note that, if required, the Verifit will
pause during the test to request that the gain control be adjusted to the Reference Test Setting (RTS). If
necessary, open the HIT chamber and, without moving the hearing instrument, adjust the gain control until
the large vertical arrow on the screen prompt disappears. Press <Continue> to resume the test.
During the AGC test, the RM500SL will pause to request that AGC controls or programmable parameters be set
for maximum AGC effect. If the manufacturer does not provide settings to produce maximum AGC effect, use
43
Vista de pagina 42
1 2 ... 38 39 40 41 42 43 44 45 46 47 48 ... 129 130

Comentarios a estos manuales

Sin comentarios

Philips 017266693 manuals

Owner’s manuals and user’s guides for Outdoor lighting Philips 017266693.
We providing 1 pdf manuals Philips 017266693 for download free by document types: Datasheet






More products and manuals for Outdoor lighting Philips

Models Document Type
14653046 Datasheet   Philips 01465/30/46, 2 pages
170864746 Datasheet   Philips Wall light 17086/47/46, 2 pages
170144786 Datasheet   Philips Wall light 17014/47/86, 2 pages
320103146 Datasheet   Philips Wall light 32010/31/46, 2 pages
168068776 Datasheet   Philips Wall light 16806/87/76, 2 pages
171308746 Datasheet   Philips Wall light 17130/87/46, 2 pages
168094776 Datasheet   Philips Wall light 16809/47/76, 2 pages
171734746 Datasheet   Philips Wall light 17173/47/46, 2 pages
017266631 Datasheet   Philips 017266631, 2 pages
171733046 Datasheet   Philips Wall light 17173/30/46, 2 pages
171059346 Datasheet   Philips Wall light 17105/93/46, 2 pages
171309346 Datasheet   Philips Wall light 17130/93/46, 2 pages
170149386 Datasheet   Philips Wall light 17014/93/86, 2 pages
170143186 Datasheet   Philips Wall light 17014/31/86, 2 pages
17264693 Datasheet   Philips 01726/46/93, 2 pages
171504716 Datasheet   Philips myGarden Wall light 17150/47/16, 2 pages
171058746 Datasheet   Philips Wall light 17105/87/46, 2 pages
016520616 Datasheet   Philips myGarden Wall light 01652/06/16, 2 pages
320109346 Datasheet   Philips Wall light 32010/93/46, 2 pages
163209316 Datasheet   Philips Ledino Wall light 16320/93/16, 2 pages