Making Measurements Using Spectrum Analyzer Mode
Using the Time-Gated Spectrum Analyzer Capability Without the Gate Utility (Option 107 only)
2. Set the sweep time to be 401 times greater than the pulse repetition
interval. For this example, the pulse repetition interval is 280
ps,
so the
sweep time is set to greater than 401 times 280
ps,
or 0.112 s. For this
example, we are using a sweep time of 120 milliseconds. Press
@EETJ
120
@.
SWEEP
CONT
SGL
GATE
ON
OFF
GATE
MENU
CENTER 50.000 MHz
SPAN 5,000 MHz
RES BW
30
kHz UBW
36
kHZ
XSWP
120
m5cc
RT
Figure 5-21. Setting the Sweep Time
5-44