I
-
I
-
GATE
DELAY
GATE
LENGTH
Spectrum Analyzer Mode Key Descriptions
Spectrum Analyzer Mode Functions
Option
107
on Ly.
sets the duration of the delay after an edge trigger before
the gate switch closes. The gate delay can be set from 1
,us
to 65.535 ms in 1 ps steps.
If the GATE DELAY softkey is accessed outside of the time
gate utility, it turns off the gate markers but does not affect
the gate utility settings. When the gate utility is entered,
the current value of the gate delay will be used. The
GATE DELAY softkey can be accessed from within the gate
utility under the Define Gate menu.
Front-Panel Key Access
CSWEEP)
Option
107
only.
sets the duration of the gate. The gate length can be set
from 1
,M
to 65.535 ms in 1 ps steps.
If the GATE LENGTH softkey is accessed outside of the time
gate utility it turns off the gate markers but does not affect
the gate utility settings. When the gate utility is entered,
the current value of the gate length will be used. The
GATE LENGTH softkey can be accessed from within the gate
utility under the Define
Gate
menu.
Front-Panel Key Access
(SWEEP)
6-77