Preparing For Use
Turning on the Analyzer for the First Time
1. To calibrate the tracking generator, connect the tracking generator output
(RF OUT 75
a2)
to the cable TV analyzer INPUT 75
D
connector, using an
appropriate cable.
NOTE
A low-loss cable should be used for accurate calibration. Use the 75
0
cable shipped with the cable
TV analyzer.
2. Press the following cable TV analyzer keys:
ICAL),
More 1 of 4 ,
More 2 of 4 , then CAL
TRK
GEM TG SIGNAL NOT FOUND will be
displayed if the tracking generator output is not connected to the cable TV
analyzer input.
3. To store this data in the area of cable TV analyzer memory that is stored
when the cable TV analyzer is turned off, press CAL STORE
When the self-calibration routines have been completed successfully, the cable
TV analyzer is ready for normal operation.