8.6
ANSI
test results
FastFacts 8.6: ANSI test results
In 1996 tests: Full-on average gain j is measured at 50 dB SPL, average gain at
RTP k is usually (but not always) measured at 60 dB SPL and the frequency
response l is run at 60 dB SPL for linear aids or at 50 dB SPL for AGC aids.
In 2003 tests: Full-on average gain is measured with a 50 dB input SPL, average
gain at RTS is always measured at 60 dB input SPL and the frequency response
curve is always run at 60 dB input SPL.