0611 RM500SL Users Guide Version 2.8 Page 65
16.3
Insertion gain in HL
FastFacts 16.3: Insertion gain in HL
See 10: Real-Ear Measurement Setup and 11: REM Screen Setup
1) Press <Tests>, then highlight & <PICK> [Insertion gain] from the REM tests list.
2) Highlight and <PICK> [Instrument] to chose from BTE, ITE, ITC, CIC, Body.
This changes the CORFIG used for 2cc targets.
3) Highlight and <PICK> fitting method to chose from FIG6, NAL-NL1, NAL-R,
POGO II, BERGER, LIBBY, USER, NONE.
An estimated aided threshold curve is calculated by shifting the unaided threshold
curve upward by the REIG. This estimate assumes a) the unaided threshold is a
sound-field threshold and b) a linear aid (i.e. REIG doesnt depend on level).
The REUR curve uses the right-hand SPL scale and is shown for reference only.