Manual Calibration
Data Explorer
Software Users Guide 5-9
5
Figure 5-2 Manual Mass Calibration Dialog Box
4. Select a calibration reference file. For information on
creating a reference file, see Section 5.3.3, Creating or
Modifying a Calibration Reference File (.REF).
5. Enter reference matching and calibration criteria:
Minimum IntensityPeaks must be above this
intensity to be considered a match. Select the unit
for Minimum Intensity, % Relative Intensity or
Absolute Counts.
Mass TolerancePeaks must be within this
tolerance of the theoretical m/z to be considered a
match. Select the unit, m/z or ppm.