Appendix C Data Explorer Toolbox (Visual Basic Macros)
C-16 Applied Biosystems
C
3. Select the mode for the analysis:
Use the entire mass rangeCalculates
average molecular weights using all peak
intensities within the X Display Range. It does
not distinguish between different polymeric
species that may be present in the mass
range.
Enter an adduct mass.
Figure 3-14 Polymer Analysis Toolbox
NOTE: If the predominant ion series is M+Na+, enter 23 for
the adduct ion. If the predominant ion series is M+Ag+,
enter 108 for the adduct ion. If you are analyzing an
unknown or a mixture, leave this field blank; the calculation
is performed without subtracting the adduct ion mass.