Adjusting the Baseline
Data Explorer
Software Users Guide 5-55
5
Troubleshooting If the baseline rises preceding and following a peak after the
correction (a hump under the peak), adjust the following
parameters in the order listed:
Decrease Flexibility
Decrease Degree
Decrease Peak Width
NOTE: Lower Peak Width values increase the time needed
for processing.
Returning to the
original spectrum
To return to the original trace, see Returning to the original
spectrum on page 5-3.