HP Data Explorer 4 Series Manual de usuario Pagina 321

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 447
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 320
Mariner Data Examples
Data Explorer
Software Users Guide 7-5
7
The combined spectrum is displayed (Figure 7-3), with
two intense peaks at 410 Da and 723 Da. Generate
extracted ion chromatograms as described below to
determine if these peaks are the coeluting components.
NOTE: If these peaks do not yield chromatograms with
profiles that correspond to the unresolved peaks, try
another spectral peak.
Figure 7-3 Combined Spectrum for Unresolved Peaks in Cytochrome C
Creating
extracted ion
chromatograms
To create extracted ion chromatograms:
1. With the Chromatogram window activated, click in
the toolbar two times to add two traces. Click an added
trace to activate it.
2. In the Spectrum window, right-click-drag over the first
peak.
The extracted ion chromatogram is displayed.
3. In the Chromatogram window, click the second added
trace to activate it.
4. In the Spectrum window, right-click-drag over the second
peak.
The extracted ion chromatogram is displayed.
Figure 7-4 illustrates the peaks deconvoluted from the
original unresolved peaks.
Vista de pagina 320
1 2 ... 316 317 318 319 320 321 322 323 324 325 326 ... 446 447

Comentarios a estos manuales

Sin comentarios